share:
delim
delim
4-Hr Shipping Facility Air Sample with radiello 145

4-Hr Shipping Facility Air Sample with radiello 145

Peaks
1.Propene
2.2-Methyl-1-propene
3.Acetaldehyde
4.2-Methylbutane
5.Pentane
6.Furan
7.Methylene chloride
8.Hexane
9.Acetonitrile
10.Benzene
11.Toluene
12.Chlorobenzene
image
GC_AR1171
ColumnRtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916)
with MXT low-dead-volume connector (cat.# 20536)
SampleShipping facility air sample
Injectionon-column
Oven
Oven Temp.:40 °C (hold 7 min) to 250 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
28.8038226
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:BFB
Ionization Mode:EI
PreconcentratorMarkes Unity
  Trap 1 Settings
Type/Sorbent :radiello 145
Desorb temp.:350 °C
Desorb flow:50 mL/min
Desorb time:300 sec
  Trap 2 Settings
Type/Sorbent:Air Toxics
Cooling temp.:30 °C
Desorb temp.:310 °C
Desorb time:3 sec
InstrumentAgilent 7890B GC & 5977A MSD
NotesThe radiello 145 passive air sampler (RAD145) utilizes a stainless steel net cartridge packed with 350 mg of graphitized charcoal (Carbograph 4). Airborne volatile organic compounds (VOCs) were adsorbed to the charcoal and then thermally desorbed and analyzed by GC-MS.

Trap 1 conditions were used for radiello desorption. Trap 2 conditions were used for Unity desorption.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us