Cannabinoids Rxi-5ms
Peaks | |
---|---|
1. | Cannabidiol![]() |
2. | Δ8-Tetrahydrocannabinol![]() |
3. | Δ9-Tetrahydrocannabinol![]() |
Peaks | |
---|---|
4. | Cannabinol![]() |
5. | 11-Hydroxy-Δ9-tetrahydrocannabinol![]() |
6. | 11-Nor-Δ9-tetrahydrocannabinol carboxylic acid![]() |

GC_CF00891
Column | Rxi-5ms, 12 m, 0.20 mm ID, 0.33 µm (cat.# 13497) |
Sample | |
Diluent: | Methanol |
Conc.: | 1000 µg/mL each component |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 25:1) |
Liner: | Single taper (4mm), base deactivated w/base deactivated wool |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C to 340 °C at 20 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Ionization Mode: | EI |
Scan Range: | 100-550 amu |