share:
delim
delim
GHB (Analyzed as GBL) in Urine on Rtx-5MS
This chromatogram may contain analytes that are synonyms of your search terms.

GHB (Analyzed as GBL) in Urine on Rtx-5MS

Peaks
1.GBL
Peaks
2.AMGB (IS)
image
GC_CF01038
ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Sample300 µg/mL GHB, 50 µg/mL AMGB in urine, converted with sulfuric acid
Injection
Inj. Vol.:1000 µL headspace-syringe split (split ratio 10:1)
Liner:1 mm Split (cat.# 20972)
Inj. Temp.:200 °C
Headspace-Syringe
Sample Temp.:100 °C
Sample Equil. Time:10 min
Oven
Oven Temp.:50 °C (hold 3 min) to 150 °C at 20 °C/min (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:20-150 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Gamma-hydroxybutyrate

,

GHB

,

gamma-butyrolactone

,

GBL

,

date rape

,

alpha-methylene-gamma-butyrola

,

AMGB

,

sexual assault

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us