share:
delim
delim
Comparison of Converted and Unconverted GHB on Rtx-BAC1

Comparison of Converted and Unconverted GHB on Rtx-BAC1

Peaks
1.GBL
Peaks
2.AMGB (IS)
image
GC_CF01039
ColumnRtx-BAC1, 30 m, 0.32 mm ID, 1.80 µm (cat.# 18003)
Sample
Conc.:50 µg/mL GHB, GBL, and AMGB (IS) in urine
A: unconverted
B: converted with sulfuric acid
Injection
Inj. Vol.:1000 µL headspace-syringe split (split ratio 10:1)
Liner:1mm Split (cat.# 20972)
Inj. Temp.:200 °C
Headspace-Syringe
Sample Temp.:100 °C
Sample Equil. Time:10 min
Oven
Oven Temp.:50 °C (hold 3 min) to 150 °C at 20 °C/min (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorFID @ 240 °C
Make-up Gas Flow Rate:40 mL/min
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Gamma-hydroxybutyrate

,

GHB

,

gamma-butyrolactone

,

GBL

,

date rape

,

date rape drug

,

alpha-methylene-gamma-butyrola

,

AMGB

,

sexual assault

,

drug-facilitated

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us