share:
delim
delim
Benzodiazepines (15 μg/mL) on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Benzodiazepines (15 μg/mL) on Rxi-5Sil MS

Peaks
1.Oxazepam
2.Lorazepam
3.Diazepam
4.Flunitrazepam
5.Prazepam
6.Clonazepam
7.Alprazolam
image
GC_CF1148
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Butyl chloride
Conc.:15 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5 mm taper splitless w/wool (cat.# 22286-200.1)
Inj. Temp.:280 °C
Purge Flow:32.2 (20:1 split) mL/min
Oven
Oven Temp.:200 °C to 330 °C at 15 °C/min (hold 3 min)
Carrier GasHe, constant linear velocity
Linear Velocity:50 cm/sec, 23.7 psi, 163.4 kPa @ 200 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:4 min
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:50-350 amu
Scan Rate:5 scans/sec
InstrumentShimadzu 2010 GC & QP2010+ MS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Benzodiazepines

,

GCMS

,

Rxi-5MS

,

Rxi-5

,

Rxi-5SilMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us