share:
delim
delim
GHB Analysis (GBL and AMGBL) on Rtx-BAC Plus 1

GHB Analysis (GBL and AMGBL) on Rtx-BAC Plus 1

Peaks
1.GBL
Peaks
2.AMGBL
image
GC_CF1164
ColumnRtx-BAC Plus 1, 30 m, 0.32 mm ID, 1.8 µm (cat.# 18004)
SampleGamma-butyrolactone (GBL) (cat.# 34077)
Alpha-methylene-gamma-butyrolactone (AMGBL) (cat.# 34079)
Injectionheadspace-loop split (split ratio 10:1)
Liner:1 mm straight inlet liner (cat.# 20972)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT3
Inj. Time:1 min
Transfer Line Temp.:125 °C
Valve Oven Temp.:125 °C
Sample Temp.:100 °C
Sample Equil. Time:10 min
Vial Pressure:30 psi
Pressurize Time:2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:80 °C (hold 0 min) to 180 °C at 10 °C/min (hold 0 min)
Carrier GasHe, constant flow
Linear Velocity:48 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesSample was prepared by injecting 1 µL of each standard into a capped headspace vial. The mass of each compound inside the headspace vial was 1 µg.

Headspace concentrator courtesy of Teledyne Tekmar, Mason, OH.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

DFSA

,

drug facilitated sexual assault

,

GHB

,

BAC

,

blood alcohol

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us