share:
delim
delim
Cresylic Acids on Stabilwax-DA

Cresylic Acids on Stabilwax-DA

Peaks
1.2,6-Dimethylphenol
2.o-Cresol
3.Phenol
4.2,4,6-Trimethylphenol
5.2,3,6-Trimethylphenol
6.o-Ethylphenol
7.2,5-Dimethylphenol
8.p-Cresol
9.2,4-Dimethylphenol
Peaks
10.m-Cresol
11.C3 Phenol
12.2,3-Xylenol
13.C3 Phenol
14.C3 Phenol
15.C3 Phenol
16.p-Ethylphenol
17.3,5-Xylenol
18.m-Ethylphenol
19.3,4-Xylenol
image
GC_CH00277
ColumnStabilwax-DA, 30 m, 0.25 mm ID, 0.50 µm (cat.# 11038)
Injectionsplit (split ratio 40:1)
Inj. Temp.:250 °C
Oven
Oven Temp.:180 °C (hold 2 min) to 260 °C at 2 °C/min
Carrier GasH2, constant flow
Flow Rate:3.2 mL/min
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
NotesWet needle split injection of cresylic acids

FID sensitivity: 64 x 10-11 AFS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cresylic acids

,

solvents

,

chemicals

,

stabilwax-da

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us