share:
delim
delim
Fatty Acids (Free) on Rtx-1
This chromatogram may contain analytes that are synonyms of your search terms.

Fatty Acids (Free) on Rtx-1

Peaks
1.Acetic acid
2.Propionic acid
3.Isobutyric acid
4.n-Butyric acid
5.Isovaleric acid
6.n-Valeric acid
7.Isocaproic acid
8.Caproic acid
9.Heptanoic acid
image
GC_CH00281
ColumnRtx -1, 30 m, 0.53 mm ID, 5.00 µm (cat.# 10179)
Samplefree fatty acid standard
Diluent:water
Conc.:10-20 ng/µL each analyte
Injection
Inj. Vol.:0.2 µL Direct
Liner:Drilled Uniliner (hole near top) (cat.# 21054)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C to 180 °C at 15 °C/min
Carrier GasH2, constant flow
Flow Rate:6 mL/min
Linear Velocity:50 cm/sec
DetectorFID @ 250 °C
NotesFID sensitivity: 4 x 10-11 AFS
Download PDF