share:

Alcohols Rtx-5

Peaks
1.Methanol
2.Ethanol
3.Acetone
4.Isopropanol
5.1-Propanol
6.Ethyl Acetate
7.2-Methyl-1-propanol
8.1-Butanol
9.3-Pentanol
10.3-Methyl-1-butanol
11.2-Methyl-1-butanol
image
GC_CH00292
ColumnRtx-5, 60 m, 0.32 mm ID, 1.00 µm (cat.# 10257)
Sampleneat alcohols mix
Injection
Inj. Vol.:0.03 µL split (split ratio 40:1)
Inj. Temp.:200 °C
Oven
Oven Temp.:25 °C (hold 4 min) to 80 °C at 8 °C/min (hold 5 min)
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 200 °C
NotesFID sensitivity: 128 x 10-11 AFS