share:

Aldehydes on Stabilwax

Peaks
1.Pentane
2.Acetaldehyde
3.Dimethoxymethane (methylal)
4.Propanal
5.2-Methylpropanal
6.Methanol
7.Pentanal
8.2-Butenal (crotonaldehyde)
9.Hexanal
Peaks
10.Undecane
11.Heptanal
12.1-Pentanol
13.Octanal
14.Tetradecane
15.Nonanal
16.2-Furancarboxaldehyde
17.Decanal
18.Benzaldehyde
19.1-Octanol
image
GC_CH00315
ColumnStabilwax, 60 m, 0.53 mm ID, 1.00 µm (cat.# 10658)
Samplealdehydes
Injection
Inj. Vol.:1.0 µL split (split ratio 8:1)
Inj. Temp.:250 °C
Oven
Oven Temp.:45 °C (hold 10 min) to 250 °C at 12 °C/min (hold 20 min)
Carrier GasHe, constant flow
Linear Velocity:51 cm/sec @ 50 °C
DetectorMS
Mode:Scan
Source Temp.:250 °C
Ionization Mode:EI