share:
delim
delim
Freon 12 and Ethylene Oxide on Rtx-1

Freon 12 and Ethylene Oxide on Rtx-1

Peaks
1.Freon® 12
2.Ethylene oxide
image
GC_CH00324
ColumnRtx-1, 30 m, 0.53 mm ID, 5.00 µm (cat.# 10179)
Injection
Inj. Vol.:50 µL split (split ratio 40:1)
Inj. Temp.:290 °C
Oven
Oven Temp.:25 °C
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 290 °C
NotesFID sensitivity: 4 x 10-11 AFS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

freon

,

dichlorodifluoromethane

,

volatiles

,

CFC

,

CFCs

,

chlorofluorocarbons

,

solvents

,

chemicals

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us