share:

Glycols on Rtx-200

Peaks
1.Ethylene glycol
2.Propylene glycol
3.Diethylene glycol
4.Tetraethylene glycol
image
GC_CH00326
ColumnRtx-200, 60 m, 0.53 mm ID, 3.00 µm (cat.# 15088)
SampleGlycol
Conc.:50 ng on-column
Injection
Inj. Vol.:1.0 µL split (split ratio 19:1)
Inj. Temp.:200 °C
Oven
Oven Temp.:40 °C to 250 °C at 8 °C/min (hold 15 min)
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
NotesFID sensitivity: 6.4 x 10-11 AFS
Flow rate: 5 cc/min