Solvent Mixture #1 on Rtx-200

Solvent Mixture #1 on Rtx-200

Peaks
1.Pentane
2.Methylene Chloride
3.Ethylene glycol
4.Heptane
5.Cyclopentanol
6.3-Hexanol
7.Acetamide
8.2-Methyl-1-pentanol
9.Furfuryl alcohol
10.Butyl ether
11.Nonane
12.Cumene
13.Ethyl amyl ketone
14.Heptanol
15.Butyl butanoate
16.Unknown
17.Benzyl Alcohol
18.Dipropylene glycol
19.benzene, diethyl-
20.Unknown
21.Unknown
22.Hexachloroethane
23.Undecane
24.1-Nonanol
25.p-Methoxyphenol
26.Triethylene glycol
27.Dodecane
28.Undecanal
29.Tridecane
30.Unknown
31.Dodecanal
32.Dicyclohexylamine
33.bis(2,2-methoxy)ethyl ether
34.Pentadecane
35.Heptadecane
36.Octadecane
37.Nonadecane
38.Eicosane
39.Acetyl tributyl citrate
40.2-Buten-1-ol
41.Formamide
42.3-Pentanol
43.1-Nitropropane
44.Dimethylformamide
45.2-Methyl-3-pentanol
46.Toluene
47.Ethyl chloroacetate
48.Dimethylacetamide
49.p-Xylene
50.sec-tetrachloroethane
51.Benzaldehyde
Peaks
52.o-Chlorotoluene
53.2,6-Dimethyl-4-heptanone
54.2-Octanone
55.o-Cresol
56.α-Methylbenzyl alcohol
57.5-Nonanone
58.Nonanal
59.Decanal
60.Unknown
61.1-Decanol
62.1-Undecanol
63.2-Dodecanone
64.1-Dodecanol
65.Tetraethylene glycol
66.Dibenzyl
67.Diethyl Phthalate
68.Tributyl phosphate
69.Diphenyl sulfone
70.Allyl alcohol
71.Unknown
72.Isopropyl acetate
73.Benzene
74.2-Nitropropane
75.Nitroethane
76.Pentanal
77.2-Bromobutane
78.1-Chloropentane
79.Cyclopentanone
80.2-Hexanol
81.Butyl acetate
82.2-Ethyl-1-butanol
83.3-Ethyl-3-pentanol
84.1,4-Dichlorobutane
85.2-Methyl-2,4-pentanediol
86.Butoxyethanol
87.1,2,3-Trichloropropane
88.1,4-Butanediol
89.Methyl hexanoate
90.1,2,4-Trimethylbenzene
91.2-Ethyl-1-hexanol
92.Dipentene/Limonene
93.Tetrahydrofurfuryl acetate
94.Unknown
95.Decahydronaphthalene
96.Unknown
97.Unknown
98.2-Decanol
99.1,2-Bis(2-methoxyethoxy)ethane
100.2-Phenoxyethanol
101.Unknown
102.Benzyl ether
image
GC_CH00336
ColumnRtx-200, 60 m, 0.53 mm ID, 3.00 µm (cat.# 15088)
SampleSolvent mix #1
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:275 °C
Split Vent Flow Rate:50 mL/min
Oven
Oven Temp.:40 °C (hold 5 min) to 285 °C at 5 °C/min
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorMS
Mode:Scan
Source Temp.:285 °C
NotesMSD: TIC mode
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

limonene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us