share:

Solvents on Stabilwax

Peaks
1.Pentane
2.Hexane
3.Methyl formate
4.Acetone
5.Tetrahydrofuran
6.1,1,1-Trichloroethane
7.Ethyl acetate
8.Methyl ethyl ketone
9.Methanol
10.Methylene chloride
Peaks
11.Chloroform
12.Benzene
13.n-Propyl alcohol
14.Toluene
15.sec-Butyl alcohol
16.n-Butyl acetate
17.Isobutyl alcohol
18.n-Butyl alcohol
19.o-Xylene
20.Ethoxyethane
image
GC_CH00345
ColumnStabilwax, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10623)
Samplesolvent mixture
Injection
Inj. Vol.:0.5 µL split (split ratio 40:1)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 4 min) to 250 °C at 5 °C/min
Carrier GasH2, constant flow
Flow Rate:3.2 mL/min
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
NotesFID sensitivity: 16 x 10-11 AFS
Download PDF