share:

Solvents on Rtx-5

Peaks
1.Acetone
2.Acrylonitrile
3.carbon disulfide impurity
4.Hexane
5.Methyl ethyl ketone
6.Ethyl acetate
7.Methyl Cellosolve®
8.Tetrahydrofuran
9.n-Butyl alcohol
Peaks
10.Isopropyl acetate
11.Heptane
12.n-Propyl acetate
13.Amyl alcohol
14.Toluene
15.n-Butyl acetate
16.n-Butyl acrylate
17.Cellosolve® acetate
18.Cyclohexanone
image
GC_CH00346
ColumnRtx-5, 60 m, 0.32 mm ID, 3.00 µm (cat.# 10287)
Samplesolvents collected on a charcoal tube and extracted in carbon disulfide
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:300 °C
Oven
Oven Temp.:50 °C (hold 4 min) to 120 °C at 5 °C/min (hold 20 min)
Carrier GasH2, constant pressure
Linear Velocity:40 cm/sec @ 50 °C
DetectorFID @ 300 °C
NotesFID sensitivity: 128 x 10-12 AFS
AcknowledgementPermission to publish this chromatogram granted by Kenda McCoy, Rexham Industrial.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Solvents

,

Rtx-5

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.