delim
delim
Styrene Impurities on Stabilwax
This chromatogram may contain analytes that are synonyms of your search terms.

Styrene Impurities on Stabilwax

Peaks
1.Butene
2.1,3-Butadiene
3.Diethylhydroxylamine
4.Acrylonitrile
5.Vinylcyclohexene
6.Toluene
7.Ethylbenzene
8.m-Xylene
9.Cumene
10.o-Xylene
11.Styrene
image
GC_CH00356
ColumnStabilwax, 30 m, 0.53 mm ID, 0.50 µm (cat.# 10640)
Sample95% pure styrene
Injection
Inj. Vol.:0.5 µL split
Inj. Temp.:150 °C
Split Vent Flow Rate:40 mL/min
Oven
Oven Temp.:40 °C (hold 10 min) to 150 °C at 10 °C/min (hold 15 min)
Carrier GasHe, constant pressure
Linear Velocity:20 cm/sec @ 40 °C
DetectorFID @ 150 °C
NotesFID sensitivity: 16 x 10-11 AFS
AcknowledgementPermission to publish this chromatogram granted by Copolymer Rubber and Chemical Corp.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Stabilwax

,

styrene

,

impurities

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us