Formaldehyde on Rt-U-BOND
This chromatogram may contain analytes that are synonyms of your search terms.

Formaldehyde on Rt-U-BOND

Peaks
1.Air
2.Carbon dioxide
3.Formaldehyde
4.Water
5.Methanol
image
GC_CH01137
ColumnRt-U-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19750)
SampleFormaldehyde (manual headspace)
Injection
Inj. Vol.:10 µL split (split ratio 10:1)
Liner:2 mm split Precision liner w/wool (cat.# 20823)
Inj. Temp.:200 °C
Split Vent Flow Rate:40 mL/min
Oven
Oven Temp.:100 °C (hold 1 min) to 150 °C at 25 °C/min (hold 3 min)
Carrier GasHe, constant pressure (7.7 psi, 53.1 kPa)
Linear Velocity:39 cm/sec @ 100 °C
DetectorTCD @ 200 °C
Make-up Gas Type:He
Data Rate:20 Hz
Sensitivity Mode:He/H2
InstrumentHP5890 GC
NotesSample Preparation: 10μl of formaldehyde solution added to a 20ml vial and heated to 60°C with CO2 added to the headspace. The formaldehyde solution is at a nominal concentration of 37% by weight with 10 to 15% methanol and the remaining amount water to stabilize the formaldehyde.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Formaldehyde

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us