share:
delim
delim
EDB/DBCP by EPA 504.1 on Rtx-CLPesticides & Rtx-CLPesticides2

EDB/DBCP by EPA 504.1 on Rtx-CLPesticides & Rtx-CLPesticides2

Peaks
1.Chloroform
2.Bromodichloromethane
3.Chlorodibromomethane
4.1,2-Dibromoethane (EDB)
Peaks
5.1,1,1,2-Tetrachloroethane
6.Bromoform
7.1,2,3-Trichloropropane
8.1,2-Dibromo-3-chloropropane (DBCP)
image
GC_EV00020
ColumnsRtx-CLPesticides 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139)
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
using IP Deactivated Guard Column 0.32 mm ID (cat.# 10044)
with Universal Angled “Y” Press-Tight Connector (cat.# 20403)
Sample
Conc.:10 pg each compound
InjectionDirect
Liner:Drilled Uniliner (hole near top) (cat.# 20335)
Inj. Temp.:200 °C
Oven
Oven Temp.:35 °C (hold 2 min) to 300 °C at 12 °C/min
Carrier GasHe, constant pressure (12 psi, 82.7 kPa)
DetectorECD @ 300 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA 504.1

,

EDB

,

DBCP

,

504.1

,

micro Extract ECD

,

EDB & DBCP

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us