share:
delim
delim
Phenols and Benzidines by U.S. EPA Methods 604 and 605 on Rtx-5

Phenols and Benzidines by U.S. EPA Methods 604 and 605 on Rtx-5

Peaks
1.Phenol
2.2-Chlorophenol
3.2-Nitrophenol
4.2,4-Dimethylphenol
5.2,4-Dichlorophenol
6.4-Chloro-3-methylphenol
Peaks
7.2,4,6-Trichlorophenol
8.2,4-Dinitrophenol
9.4-Nitrophenol
10.2-Methyl-4,6-dinitrophenol
11.Pentachlorophenol
12.Benzidine
13.3,3'-Dichlorobenzidine
image
GC_EV00032
ColumnRtx-5, 30 m, 0.32 mm ID, 1.00 µm (cat.# 10254)
Sample604 Phenols calibration mix (cat.# 31029)
605 Benzidines calibration mix (cat.# 31030)
Conc.: 1-3 ng per component on-column
Injection
Inj. Vol.:1.5 µL split (split ratio 100:1)
Inj. Temp.:310 °C
Oven
Oven Temp.:110 °C to 290 °C at 8 °C/min
Carrier GasH2, constant linear velocity
Linear Velocity:40 cm/sec
DetectorFID @ 310 °C
NotesFID sensitivity: 2 x 10-11 AFS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us