share:
delim
delim
Volatile Petroleum Hydrocarbons (VPH) Massachusetts Department of Environmental Protection on Rtx-502.2

Volatile Petroleum Hydrocarbons (VPH) Massachusetts Department of Environmental Protection on Rtx-502.2

Peaks
1.Methanol
2.n-Pentane (40ng)
3.2-Methylpentane (60ng)
4.Methyl tert-butyl ether (60ng)
5.2,2,4-Trimethylpentane (60ng)
6.Benzene (20ng)
7.Toluene
Peaks
8.n-Nonane
9.Ethylbenzene (20ng)
10.p- and m-Xylene (40ng ea.)
11.o-Xylene (40ng)
12.1,2,4-Trimethylbenzene (40ng)
13.Naphthalene (40ng)
14.2,5-Dibromotoluene (SS)
image
GC_EV00063
ColumnRtx-502.2, 105 m, 0.53 mm ID, 3.0 µm (cat.# 10910)
Sample
Conc.: on-column at levels listed
Injectionpurge and trap split
Purge and Trap
Instrument:Tekmar Model LSC 2000
Trap Type:BTEX
Purge: 11 min, flow 40 ml/min
Dry Purge: 2 min
Desorb Preheat Temp:245 °C
Desorb: 2 min @ 250 °C
Bake: 6 min @ 260 °C
Oven
Oven Temp.:45 °C to 90 °C at 3 °C/min to 140 °C at 5 °C/min to 230 °C at 45 °C/min (hold 8 min)
Carrier GasHe, constant flow
Flow Rate:15 mL/min
DetectorPID and FID
NotesFor VPH analysis on an Rtx-502.2 column, use PID for aromatic
compounds and FID for aliphatic compounds
AcknowledgementChromatograms courtesy of Severn Trent Laboratories, Burlington, VT.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

volatile

,

hydrocarbon

,

xylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us