Volatile Petroleum Hydrocarbons (VPH) Massachusetts Department of Environmental Protection on Rtx-502.2
Peaks | |
---|---|
1. | Methanol![]() |
2. | n-Pentane (40ng)![]() |
3. | 2-Methylpentane (60ng)![]() |
4. | Methyl tert-butyl ether (60ng)![]() |
5. | 2,2,4-Trimethylpentane (60ng)![]() |
6. | Benzene (20ng)![]() |
7. | Toluene![]() |
Peaks | |
---|---|
8. | n-Nonane![]() |
9. | Ethylbenzene (20ng)![]() |
10. | p- and m-Xylene (40ng ea.) |
11. | o-Xylene (40ng)![]() |
12. | 1,2,4-Trimethylbenzene (40ng)![]() |
13. | Naphthalene (40ng)![]() |
14. | 2,5-Dibromotoluene (SS)![]() |

GC_EV00063
Column | Rtx-502.2, 105 m, 0.53 mm ID, 3.0 µm (cat.# 10910) |
Sample | |
Conc.: | on-column at levels listed |
Injection | purge and trap split |
Purge and Trap | |
Instrument: | Tekmar Model LSC 2000 |
Trap Type: | BTEX |
Purge: | 11 min, flow 40 ml/min |
Dry Purge: | 2 min |
Desorb Preheat Temp: | 245 °C |
Desorb: | 2 min @ 250 °C |
Bake: | 6 min @ 260 °C |
Oven | |
Oven Temp.: | 45 °C to 90 °C at 3 °C/min to 140 °C at 5 °C/min to 230 °C at 45 °C/min (hold 8 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 15 mL/min |
Detector | PID and FID |
Notes | For VPH analysis on an Rtx-502.2 column, use PID for aromatic compounds and FID for aliphatic compounds |
Acknowledgement | Chromatograms courtesy of Severn Trent Laboratories, Burlington, VT. |