share:

Diesel Fuel on Rtx-5

image
GC_EV00070
ColumnRtx-5, 30 m, 0.53 mm ID, 1.00 µm (cat.# 10255)
Samplediesel fuel #2 composite (cat.# 31093)
Conc.:500 ng/µL
InjectionDirect
Inj. Temp.:280 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 310 °C at 12 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:65 cm/sec
DetectorFID @ 310 °C
NotesFID sensitivity: 8 x 10-11 AFS
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us