share:
delim
delim
Wisconsin GRO Mix, plus C12 on Rtx-502.2

Wisconsin GRO Mix, plus C12 on Rtx-502.2

Peaks
1.3-Methylpentane
2.2,2,4-Trimethylpentane (isooctane)
3.Benzene
4.Toluene
5.Ethylbenzene
Peaks
6.m-Xylene
7.o-Xylene
8.1,2,4-Trimethylbenzene
9.Dodecane
10.Naphthalene
image
GC_EV00071
ColumnRtx-502.2, 105 m, 0.53 mm ID, 3.00 µm (cat.# 10910)
SampleGRO Mix (cat.# 30069)
Conc.:200 ppb each in 5 mL water
Injectionpurge and trap splitless
Inj. Temp.:200 °C
Purge and Trap
Trap Type:Tenax, silica gel, charcoal
Purge: 12 min, flow 40 mL/min
Desorb Preheat Temp.:175 °C
Desorb: 2 min @ 180 °C, flow 10 mL/min
Oven
Oven Temp.:40 °C (hold 1 min) to 100 °C at 5 °C/min to 240 °C at 8 °C/min (hold 8 min)
Carrier GasHe, constant flow
Flow Rate:10 mL/min
DetectorFID @ 250 °C
NotesFID Sensitivity: 16 x 10-11 AFS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Wisconsin

,

GRO

,

RTx-502.2

,

purge and trap

,

P&T

,

purge

,

trap

,

xylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us