Glycols on Rtx-BAC1 & Rtx-BAC2
Peaks | |
---|---|
1. | Ethylene glycol![]() |
2. | Propylene glycol![]() |

GC_EV00474
Columns | Rtx-BAC1 30 m, 0.32 mm ID, 1.80 µm (cat.# 18003) |
and Rtx-BAC2 30 m, 0.32 mm ID, 1.20 µm (cat.# 18002) | |
Sample | |
Conc.: | 1% Glycols in methanol |
Injection | |
Inj. Vol.: | 0.5 µL split (split ratio 100:1) |
Liner: | Agilent 4mm single gooseneck |
Inj. Temp.: | 240 °C |
Oven | |
Oven Temp.: | 100 °C to 240 °C at 5 °C/min (hold 5 min) |
Carrier Gas | He, constant pressure (12 psi, 82.7 kPa) |
Linear Velocity: | 37 cm/sec |
Detector | FID @ 240 °C |
Notes | Septa Purge: 5.0 cc/min. Column Flow Rate: 2.1 mL/min. Make-up gas flow: 40 cc/min. |