Polynuclear Aromatic Hydrocarbons US EPA Method 610 on Rtx-440
Peaks | |
---|---|
1. | Naphthalene![]() |
2. | Acenaphthylene![]() |
3. | Acenaphthene![]() |
4. | Fluorene![]() |
5. | Phenanthrene![]() |
6. | Anthracene![]() |
7. | Fluoranthene![]() |
8. | Pyrene![]() |
Peaks | |
---|---|
9. | Benzo[a]anthracene![]() |
10. | Chrysene![]() |
11. | Benzo[b]fluoranthene![]() |
12. | Benzo[k]fluoranthene![]() |
13. | Benzo[a]pyrene![]() |
14. | Indeno[1,2,3-cd]pyrene![]() |
15. | Dibenz[a,h]anthracene![]() |
16. | Benzo[ghi]perylene![]() |

GC_EV00732
Column | Rtx-440, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12923) |
Sample | SV Calibration Mix #5 / 610 PAH Mix (cat.# 31011) |
Diluent: | Methylene chloride |
Conc.: | 20 ppm each compound |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.4 min) |
Liner: | 4mm Splitless (cat.# 20772) |
Inj. Temp.: | 320 °C |
Oven | |
Oven Temp.: | 40 °C (hold 2 min) to 240 °C at 30 °C/min to 320 °C at 8 °C/min (hold 5 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 3.6 mL/min |
Detector | FID @ 320 °C |