Chlorinated Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides by EPA 551.1 on Rtx-1301

GC_EV00758
Column | Rtx-1301, 30 m, 0.25 mm ID, 1.00 µm (cat.# 16053) |
Sample | Disinfection Byproducts and Chlorinated Solvents Mix (cat.# 30615) |
Method 551.1 Pesticide/Herbicide Mix (cat.# 32438) | |
Conc.: | 5-10 µg/mL each analyte |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.5 min) |
Liner: | 4mm Split w/Wool (cat.# 20781) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 35 °C (hold 22 min) to 145 °C at 10 °C/min (hold 2 min) to 225 °C at 20 °C/min (hold 15 min) to 260 °C at 10 °C/min (hold 30 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 30 cm/sec @ 35 °C |
Detector | ECD @ 290 °C |