share:
This chromatogram may contain analytes that are synonyms of your search terms.

Aroclor PCBs on Rtx-PCB

image
GC_EV00767
ColumnRtx-PCB, 60 m, 0.25 mm ID, 0.25 µm (cat.# 13226)
SampleAroclor 1242 (cat.# 32009)
Aroclor 1254 (cat.# 32011)
Aroclor 1262 (cat.# 32409)
Conc.:333 ppm each
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:Splitless taper (4 mm) w/wool (cat.# 22405)
Inj. Temp.:280 °C
Oven
Oven Temp.:100 °C (hold 1 min) to 200 °C at 30 °C/min to 320 °C at 2 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.1 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:50-550 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Aroclors

,

PCB

,

polychlorinated biphenyls

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.