share:
delim
delim
Organochlorine Pesticides US EPA Method 8081 on Rtx-440

Organochlorine Pesticides US EPA Method 8081 on Rtx-440

Peaks
1.2,4,5,6-Tetrachloro-m-xylene (surr.)
2.α-BHC
3.γ-BHC
4.β-BHC
5.δ-BHC
6.Heptachlor
7.Aldrin
8.Heptachlor epoxide
9.γ-Chlordane
10.α-Chlordane
11.Endosulfan I
Peaks
12.4,4'-DDE
13.Dieldrin
14.Endrin
15.4,4'-DDD
16.Endosulfan II
17.Endrin aldehyde
18.4,4'-DDT
19.Endosulfan sulfate
20.Methoxychlor
21.Endrin ketone
22.Decachlorobiphenyl (surr.)
image
GC_EV00776
ColumnRtx-440, 30 m, 0.32 mm ID, 0.50 µm (cat.# 12939)
using Siltek Guard Column 5 m, 0.32 mm ID (cat.# 10027)
with SeCure “Y” Connector Kit (cat.# 20277)
SampleOrganochlorine Pesticide Mix AB #2 (cat.# 32292)
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000)
Conc.: Organochlorine Pesticide Mix AB #2, 8-80µg/mL each component in hexane/toluene (cat.# 32292), Pesticide Surrogate Mix, 200µg/mL each component in acetone (cat.# 32000)
Injection
Inj. Vol.:0.5 µL splitless (hold 0.75 min)
Liner:3.5 mm Single Gooseneck (cat.# 20961)
Inj. Temp.:250 °C
Oven
Oven Temp.:110 °C (hold 0.5 min) to 230 °C at 25 °C/min to 330 °C at 15 °C/min (hold 1 min)
Carrier GasH2, constant pressure
Linear Velocity:60 cm/sec @ 300 °C
DetectorECD @ 330 °C
InstrumentShimadzu 2010 GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us