share:
delim
delim
Volatile Organics (Canada) on Rtx-VMS
This chromatogram may contain analytes that are synonyms of your search terms.

Volatile Organics (Canada) on Rtx-VMS

Peaks
1.1,1-Dichloroethene
2.Methylene Chloride
3.Chloroform
4.Carbon Tetrachloride
5.Benzene
6.1,2-Dichloroethane
7.Trichloroethylene
8.Bromodichloromethane
9.Toluene
Peaks
10.Tetrachloroethylene
11.Dibromochloromethane
12.Chlorobenzene
13.Ethylbenzene
14.m-Xylene
15.p-Xylene
16.o-Xylene
17.Bromoform
18.1,4-Dichlorobenzene
19.1,2-Dichlorobenzene
image
GC_EV00813
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
SampleCanadian Drinking Water Volatiles Mix (2000μg/mL each component) (cat.# 30610)
Diluent:25mL water
Conc.: 50 ppb each analyte
Injectionpurge and trap split (split ratio 35:1)
Liner:1mm Split (cat.# 20972-214.1)
Inj. Temp.:200 °C
Purge and Trap
Instrument:OI 4660 Eclipse Purge and Trap
Trap Type:#10 (Tenax/silica gel/carbon molecular sieve)
Purge: 11 min, flow 40 mL/min
Desorb Preheat Temp.:185 °C
Desorb: 0.5 min @ 190 °C, flow 35.0 mL/min
Bake: 6 min @ 210 °C
Transfer Line Temp.:150 °C
Oven
Oven Temp.:35 °C (hold 5 min) to 70 °C at 5 °C/min to 220 °C at 20 °C/min (hold 3 min)
Carrier GasHe, constant linear velocity
Linear Velocity:30 cm/sec @ 35 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:150 °C
Ionization Mode:EI
Scan Range:35-250 amu
NotesSample Temp.: 40 °C
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us