share:
delim
delim
Polycyclic Aromatic Hydrocarbons by EPA Method 610 on Rxi-5ms

Polycyclic Aromatic Hydrocarbons by EPA Method 610 on Rxi-5ms

Peaks
1.Naphthalene
2.Acenaphthylene
3.Acenaphthene
4.Fluorene
5.Phenanthrene
6.Anthracene
7.Fluoranthene
8.Pyrene
Peaks
9.Benz[a]anthracene
10.Chrysene
11.Benzo[b]fluoranthene
12.Benzo[k]fluoranthene
13.Benzo[a]pyrene
14.Indeno[1,2,3-cd]pyrene
15.Dibenz[a,h]anthracene
16.Benzo[ghi]perylene
image
GC_EV00824
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleSV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
Conc.:10 ppm each analyte (10 ng on-column)
Injection
Inj. Vol.:1.0 µL splitless (hold 0.1 min)
Liner:Drilled Uniliner (hole near bottom) (cat.# 20771)
Inj. Temp.:275 °C
Oven
Oven Temp.:75 °C (hold 0.5 min) to 245 °C at 25 °C/min to 330 °C at 4 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Solvent Delay Time:2 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:35-550 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

610

,

PAHs

,

polynuclear aromatic hydrocarbons

,

polycyclic aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us