Polycyclic Aromatic Hydrocarbons by EPA Method 610 on Rxi-5ms
Peaks | |
---|---|
1. | Naphthalene![]() |
2. | Acenaphthylene![]() |
3. | Acenaphthene![]() |
4. | Fluorene![]() |
5. | Phenanthrene![]() |
6. | Anthracene![]() |
7. | Fluoranthene![]() |
8. | Pyrene![]() |
Peaks | |
---|---|
9. | Benz[a]anthracene![]() |
10. | Chrysene![]() |
11. | Benzo[b]fluoranthene![]() |
12. | Benzo[k]fluoranthene![]() |
13. | Benzo[a]pyrene![]() |
14. | Indeno[1,2,3-cd]pyrene![]() |
15. | Dibenz[a,h]anthracene![]() |
16. | Benzo[ghi]perylene![]() |

GC_EV00824
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
Sample | SV Calibration Mix #5 / 610 PAH Mix (cat.# 31011) |
Conc.: | 10 ppm each analyte (10 ng on-column) |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.1 min) |
Liner: | Drilled Uniliner (hole near bottom) (cat.# 20771) |
Inj. Temp.: | 275 °C |
Oven | |
Oven Temp.: | 75 °C (hold 0.5 min) to 245 °C at 25 °C/min to 330 °C at 4 °C/min (hold 1 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Solvent Delay Time: | 2 min |
Tune Type: | DFTPP |
Ionization Mode: | EI |
Scan Range: | 35-550 amu |
Instrument | HP6890 GC & 5973 MSD |