1,4-Dioxane, 2,4-Dinitrophenol, and Pyridine US EPA Method 8270D on Rxi-1ms
This chromatogram may contain analytes that are synonyms of your search terms.

1,4-Dioxane, 2,4-Dinitrophenol, and Pyridine US EPA Method 8270D on Rxi-1ms

Peaks
1.1,4-Dioxane
2.2,4-Dinitrophenol
3.Pyridine
image
GC_EV00830
ColumnRxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323)
Sample8270 MegaMix (cat.# 31850)
Benzoic acid (cat.# 31879)
8270 Benzidines Mix (cat.# 31852)
Acid Surrogate Mix (4/89 SOW) (cat.# 31025)
Revised B/N Surrogate Mix (cat.# 31887)
1,4-Dioxane (cat.# 31853)
Conc.:0.5 µg/mL each analyte (0.5ng on column)
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Drilled Uniliner (hole near bottom) (cat.# 20771)
Inj. Temp.:250 °C
Oven
Oven Temp.:50 °C (hold 0.5 min) to 245 °C at 25 °C/min to 330 °C at 6 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Solvent Delay Time:2 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:35-550 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us