share:
delim
delim
An Rxi-5ms column provides the most symmetric peak for the basic compound pyridine

An Rxi-5ms column provides the most symmetric peak for the basic compound pyridine

image
GC_EV00857
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
Sample2 ng pyridine on-column
Conc.:40 µg/mL
Injection
Inj. Vol.:1.0 µL split (split ratio 20:1)
Liner:4mm Split w/Wool (cat.# 20781)
Inj. Temp.:250 °C
Oven
Oven Temp.:50 °C (hold 3 min) to 180 °C at 35 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesColumns tested:
DB-5, DB-5ms, HP-5ms, Rxi-5ms,
VF-5ms, ZB-5, ZB-5ms
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

peak shape

,

inertness

,

competitor comparison

,

pyridine

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us