share:
delim
delim
Organochlorine Pesticides on Rtx-CLPesticides2
This chromatogram may contain analytes that are synonyms of your search terms.

Organochlorine Pesticides on Rtx-CLPesticides2

Peaks
1.2,4,5,6-Tetrachloro-m-xylene (SS)
2.α-BHC
3.γ-BHC
4.β-BHC
5.δ-BHC
6.Heptachlor
7.Aldrin
8.Heptachlor epoxide
9.trans-Chlordane*
10.cis-Chlordane*
11.Endosulfan I
Peaks
12.4,4'-DDE
13.Dieldrin
14.Endrin
15.4,4'-DDD
16.Endosulfan II
17.4,4'-DDT
18.Endrin aldehyde
19.Endosulfan sulfate
20.Methoxychlor
21.Endrin ketone
22.Decachlorobiphenyl (SS)
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV00895
ColumnRtx-CLPesticides2, 20 m, 0.18 mm ID, 0.14 µm (cat.# 42302)
SampleOrganochlorine Pesticide Mix AB #2 (8-80µg/mL each component in hexane/toluene 1:1) (cat.# 32292)
Pesticide Surrogate Mix, EPA 8080, 8081 (200µg/mL each component in acetone) (cat.# 32000)
Injection
Inj. Vol.:1.0 µL split (split ratio 20:1)
Liner:Siltek Gooseneck Splitless (4mm) (cat.# 20798-214.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:275 °C
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorECD @ 250 °C
NotesColumn Heater: Gerstel Modular Accelerated Column Heater
Gerstel module temp.: 140°C to 220°C (hold 1 min.) @ 120°C/min., to 305°C @ 50°C/min., to 330°C @ 300°C/min. (hold 2 min.)

This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual-columns connected to a 5-meter guard column using a “Y” Press-Tight® connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

CLP

,

CLP2

,

chlorinated pesticides

,

Gerstel

,

MACH

,

Gerstel MACH

,

OCP

,

pesticides

,

organochlorine pesticides

,

BHC

,

Chlordane

,

Endosulfan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us