share:
delim
delim
Polycyclic Aromatic Hydrocarbons on Rxi-17 column
This chromatogram may contain analytes that are synonyms of your search terms.

Polycyclic Aromatic Hydrocarbons on Rxi-17 column

PeakstR (min)
1.Naphthalene4.70
2.1-Methylnaphthalene5.28
3.2-Methylnapthalene5.46
4.Acenaphthylene6.45
5.Acenaphthene6.60
6.Fluorene7.18
7.Phenanthrene9.10
8.Anthracene9.14
9.Fluoranthene12.50
10.Pyrene13.33
11.Benz[a]anthracene16.32
12.Chrysene16.58
13.Benzo[b]fluoranthene19.70
PeakstR (min)
14.Benzo[k]fluoranthene19.78
15.Benzo(j)fluoranthene19.95
16.Benzo[a]pyrene21.17
17.3-Methylcholanthrene21.97
18.Dibenz(a,h)acridine24.33
19.Dibenz[a,j]acridine24.39
20.Indeno(1,2,3-cd)pyrene25.04
21.Dibenz[a,h]anthracene25.07
22.Benzo[ghi]perylene26.43
23.7H-Dibenzo(c,g)carbazole27.75
24.Dibenzo(a,E)pyrene34.46
25.Dibenzo(a,i)pyrene35.80
26.Dibenzo(a,h)pyrene36.73
image
GC_EV00925
ColumnRxi-17, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13523)
SampleSV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
PAH Supplement Mix for Method 8100 (cat.# 31857)
1-methylnaphthalene (cat.# 31283)
2-methylnaphthalene (cat.# 31285)
Conc.:20 µg/mL each component
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.2 min)
Liner:Drilled Uniliner (hole near top) (cat.# 21055)
Inj. Temp.:300 °C
Pulse Pressure:20 psi (137.9kPa)
Pulse Time:0.3 min
Purge Flow:40 mL/min
Oven
Oven Temp.:90 °C (hold 1.0 min) to 215 °C at 25 °C/min (hold 0.5 min) to 235 °C at 4 °C/min (hold 0 min) to 280 °C at 15 °C/min (hold 0 min) to 320 °C at 4 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Analyzer Type:Quadrupole
Solvent Delay Time:4.0 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:50-550 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAHs

,

polyaromatic hydrocarbons

,

polycyclic aromatic hydrocarbons

,

polynuclear aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us