share:
delim
delim
Organochlorine Pesticide Mix AB #2 Rtx-CLPesticides and Rtx-CLPesticides2 (0.25mm ID column set)

Organochlorine Pesticide Mix AB #2 Rtx-CLPesticides and Rtx-CLPesticides2 (0.25mm ID column set)

Peaks
1.2,4,5,6-tetrachloro-m-xylene (SS)
2.α-BHC
3.γ-BHC
4.β-BHC
5.δ-BHC
6.Heptachlor
7.Aldrin
8.Heptachlor epoxide (isomer B)
9.trans-Chlordane*
10.cis-Chlordane*
11.Endosulfan I
Peaks
12.4,4'-DDE
13.Dieldrin
14.Endrin
15.4,4'-DDD
16.Endosulfan II
17.4,4'-DDT
18.Endrin aldehyde
19.Endosulfan sulfate
20.Methoxychlor
21.Endrin ketone
22.decachlorobiphenyl (SS)
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV00941
ColumnsRtx-CLPesticides 30 m, 0.25 mm ID, 0.25 µm (cat.# 11123)
and Rtx-CLPesticides2 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323)
using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029)
with Siltek-treated universal “Y” Press-Tight connector (cat.# 20486)
SampleOrganochlorine pesticide mix AB #2 (cat.# 32292)
Pesticide surrogate mix, EPA 8080, 8081 (cat.# 32000)
Injection
Inj. Vol.:0.5 µL splitless (hold 0.5 min)
Liner:Cyclo double taper (2 mm) (cat.# 20908)
Inj. Temp.:250 °C
Oven
Oven Temp.:125 °C to 200 °C at 45 °C/min to 230 °C at 12.5 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant linear velocity
Linear Velocity:41 cm/sec @ 125 °C
DetectorECD @ 330 °C
NotesThis chromatogram was obtained using an Agilent μ-ECD. To obtain comparable results, you will need to employ a μ-ECD in addition to dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

BHC

,

Chlordane

,

Endosulfan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us