share:
delim
delim
Polycyclic Aromatic Hydrocarbons on Rtx-50

Polycyclic Aromatic Hydrocarbons on Rtx-50

Peaks
1.Naphthalene
2.1-Methylnaphthalene
3.2-Methylnapthalene
4.Acenaphthylene
5.Acenapthene
6.Fluorene
7.Phenanthrene
8.Anthracene
9.Fluoranthene
10.Pyrene
11.Benzo(a)anthracene
12.Chrysene
13.Benzo(b)fluoranthene
Peaks
14.Benzo(k)fluoranthene
15.Benzo(j)fluoranthene
16.Benzo(a)pyrene
17.3-Methylcholanthrene
18.Dibenzo(a,h)acridine
19.Dibenzo(a,j)acridine
20.Indeno(1,2,3-cd)pyrene
21.Dibenzo(a,h)anthracene
22.Benzo(ghi)perylene
23.7H-Dibenzo(c,g)carbazole
24.Dibenzo(a,e)pyrene
25.Dibenzo(a,i)pyrene
26.Dibenzo(a,h)pyrene
image
GC_EV00942
ColumnRtx-50, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10523)
SampleSV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
PAH Supplement Mix for Method 8100 (cat.# 31857)
1-methylnaphthalene (cat.# 31283)
2-methylnaphthalene (cat.# 31285)
Conc.:20 µg/mL
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.2 min)
Liner:Drilled Uniliner (hole near top) (cat.# 21055)
Inj. Temp.:300 °C
Pulse Pressure:20 psi (137.9kPa)
Pulse Time:0.3 min
Purge Flow:40 mL/min
Oven
Oven Temp.:90 °C (hold 2.0 min) to 215 °C at 25 °C/min to 220 °C at 2.5 °C/min to 300 °C at 10 °C/min to 320 °C at 2.5 °C/min (hold 15 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Analyzer Type:Quadrupole
Solvent Delay Time:4.0 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:50-550 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAHs

,

semivolatiles

,

polycyclic aromatic hydrocarbons

,

polynuclear aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.