share:
delim
delim
Tetrachlorodibenzofuran Congeners on Rtx-Dioxin2
This chromatogram may contain analytes that are synonyms of your search terms.

Tetrachlorodibenzofuran Congeners on Rtx-Dioxin2

image
GC_EV00949
ColumnRtx-Dioxin2, 60 m, 0.25 mm ID, 0.25 µm (cat.# 10758)
SampleWMS-01 Reference Material, Wellington Laboratories
Injectionsplitless
Inj. Temp.:250 °C
Oven
Oven Temp.:130 °C (hold 1.0 min) to 200 °C at 40 °C/min to 235 °C at 3.0 °C/min to 300 °C at 5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorMicromass Ultima high-resolution mass spectrometer
Ionization Mode:EI
NotesDetector mode: SIR
AcknowledgementChromatograms courtesy of Terry Kolic, Karen MacPherson, Eric Reiner, Ontario Ministry of Environment, Toronto, Ontario, Canada
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

TCDF

,

congeners

,

HRMS

,

dioxin

,

furan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us