Polycyclic Aromatic Hydrocarbons Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Polycyclic Aromatic Hydrocarbons Rxi-5Sil MS

PeakstR (min)
1.Naphthalene4.93
2.2-Methylnaphthalene5.58
3.1-Methylnaphthalene5.68
4.2-Fluorobiphenyl (SS)5.93
5.Acenaphthylene6.45
6.Acenaphthene6.62
7.Fluorene7.12
8.Phenanthrene8.06
9.Anthracene8.11
10.Fluoranthene9.23
PeakstR (min)
11.Pyrene9.45
12.p-Terphenyl-d14 (IS)9.61
13.Benzo[a]anthracene10.65
14.Chrysene10.69
15.Benzo[b]fluoranthene11.96
16.Benzo[k]fluoranthene12.00
17.Benzo[a]pyrene12.42
18.Perylene-d12 (IS)12.51
19.Indeno[1,2,3-cd]pyrene14.19
20.Dibenzo[a,h]anthracene14.23
21.Benzo[ghi]perylene14.65
image
GC_EV00970
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SamplePAH mix, 1 µL of 0.005 µg/mL (IS 2 µg/mL)
SV calibration mix #5 / 610 PAH mix (cat.# 31011)
1-Methylnaphthalene (cat.# 31283)
2-Methylnaphthalene (cat.# 31285)
2-Fluorobiphenyl (cat.# 31091)
Conc.:5 pg on-column
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.15 min)
Liner:Drilled Uniliner (hole near top) w/wool (cat.# 21055-200.5)
Inj. Temp.:300 °C
Pulse Pressure:20 psi (137.9kPa)
Pulse Time:0.2 min
Purge Flow:60 mL/min
Oven
Oven Temp.:50 °C (hold 0.5 min) to 290 °C at 25 °C/min to 320 °C at 5 °C/min
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
Transfer Line Temp.:290 °C
Ionization Mode:EI
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Rxi-5SilMS

,

PAH

,

polycyclic aromatic hydrocarbons

,

polynuclear aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us