Herbicides on Rtx-CLPesticides & Rtx-CLPesticides2
This chromatogram may contain analytes that are synonyms of your search terms.

Herbicides on Rtx-CLPesticides & Rtx-CLPesticides2

Peaks
1.Dalapon methyl ester
2.3,5-Dichlorobenzoic acid methyl ester (SS)
3.4-Nitroanisole
4.DCAA methyl ester (SS)
5.Dicamba methyl ester
6.MCPP methyl ester
7.MCPA Methyl ester
8.Dichlorprop, methyl ester
9.4,4'-DBOB (IS)
10.2,4-D methyl ester
Peaks
11.Pentachloroanisole
12.2,4,5-TP, methyl ester
13.2,4,5-T methyl ester
14.Chloramben, methyl ester
15.2,4-DB methyl ester
16.Dinoseb methyl ester
17.Bentazon methyl ester
18.DCPA methyl ester (Chlorthal-dimethyl)
19.Pichloram methyl ester
20.Acifluorfen methyl ester
C. contaminant
image
GC_EV00971
ColumnsRtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
using Rxi deactivated guard column 5 m, 0.32 mm ID (cat.# 10039)
with universal “Y” Press-Tight connector (cat.# 20405-261)
Sample200 ng/mL herbicide mix #1 (cat.# 32055)
1,000 ng/mL dalapon methyl ester (cat.# 32057)
20,000 ng/mL herbicide mix #3 (cat.# 32059)
200 ng/mL herbicide mix #4 (cat.# 32062)
250 ng/mL 4,4'-dibromooctafluorobiphenyl (cat.# 32053)
400 ng/mL 2,4-dichlorophenyl acetic acid methyl ester (cat.# 32050)
Diluent:hexane
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:Cyclo double taper (4 mm) (cat.# 20895)
Inj. Temp.:250 °C
Oven
Oven Temp.:70 °C (hold 0.5 min) to 190 °C at 25 °C/min (hold 1 min) to 300 °C at 11 °C/min (hold 5 min)
Carrier GasHe, constant pressure
Linear Velocity:36 cm/sec @ 70 °C
DetectorMicro_ECD @ 325 °C
InstrumentAgilent/HP6890 GC
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

dalapon

,

methyl esters

,

herbicides

,

dual column

,

CLP

,

pesticides

,

8151A

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us