share:
delim
delim
Aroclor PCBs on Rtx-CLPesticides

Aroclor PCBs on Rtx-CLPesticides

Peaks
1.Decachlorobiphenyl (DCB)
image
GC_EV00993-998
ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
SamplePCB kit #1 diluted to 1,000 ppb in hexane (cat.# 32089)
Decachlorobiphenyl (BZ #209) diluted to 100 ppb in acetone (cat.# 32029)
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.3 min)
Liner:Cyclo double taper (4 mm) (cat.# 20895)
Inj. Temp.:250 °C
Pulse Pressure:30 psi (206.8kPa)
Oven
Oven Temp.:120 °C to 200 °C at 45 °C/min to 230 °C at 15 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Linear Velocity:60 cm/sec
DetectorMicro-ECD @ 330 °C
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

1016

,

1232

,

1242

,

1248

,

1254

,

1260

,

Rtx-CLP

,

Aroclor

,

1221

,

PCB

,

Rtx-CLPesticides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us