share:
delim
delim
Comparison of 2,4-Dinitrophenol response with Semivolatiles wool and MS Certified wool
This chromatogram may contain analytes that are synonyms of your search terms.

Comparison of 2,4-Dinitrophenol response with Semivolatiles wool and MS Certified wool

Peaks
1.3-Nitroaniline
2.2,4-Dinitrophenol
3.Pentachlorophenol
4.Phenanthrene
image
GC_EV01032
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SamplePhenols/Anilines/Pesticides Test Mix (cat.# 35245)
Conc.:10 ng on-column
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:Gooseneck Splitless (4mm) w/Semivolatiles Wool (cat.# 20798-231.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:50 °C (hold 3.5 min) to 180 °C at 35 °C/min (hold 3.5 min) to 300 °C at 20 °C/min (hold 0.5 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorFID @ 330 °C
NotesFID:
Hydrogen: 30mL/min.
Air: 400mL/min.
Nitrogen: 25mL/min.

10ng 2,4-dinitrophenol (DNP) on-column, single gooseneck liner with wool, FID analysis with phenanthrene internal standard.

RELATED CHROMATOGRAM LIBRARY SEARCHES

4-dinitrophenol

,

24dnp

,

24-DNP

,

response

,

wool

,

semivolatiles wool

,

SVOA wool

,

SVOC wool

,

2

,

4-DNP

,

response factor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us