Explosives by US EPA Method 8095 Rxi-5MS with PTV On-Column Liner

GC_EV01041
Column | Rxi-5ms, 6 m, 0.53 mm ID, 0.5 µm (cat.# 563153) |
using IP Deactivated Guard Column 5 m, 0.53 mm ID (cat.# 10045) | |
Sample | 8095 Calibration Mix A (cat.# 31607) |
8095 Calibration Mix B (cat.# 31608) | |
Diluent: | acetonitrile |
Injection | PTV splitless (hold 0.35 min) |
Liner: | PTV On-Column Liner (cat.# 24976) |
Inlet Temp. Program: | 55 °C to 285 °C at 10 °C/min (hold 10 min) |
PTV Splitless | |
Purge Flow: | 15 mL/min |
Oven | |
Oven Temp.: | 50 °C to 280 °C at 10 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 60 cm/sec @ 300 °C |
Detector | µ-ECD @ 300 °C |
Make-up Gas Flow Rate: | 60 mL/min |
Make-up Gas Type: | N2 |
Notes | Absolute area reproducibility improves for all compounds, and sensitive compound responses improve dramatically because of the lack of contact with the injection port. Nitroglycerin: Absolute Area % RSD = 2.6% • Relative Area % RSD = 1.6% TNT: Absolute Area % RSD = 1.5% • Relative Area % RSD = 1.4% |