Florisil Cartridge Comparison on Rtx-CLPesticides2

Florisil Cartridge Comparison on Rtx-CLPesticides2

image
GC_EV01093
ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.50 µm (cat.# 11325)
SampleFlorisil Cartridge Check Standard (2,4,5-trichlorophenol) (cat.# 32017)
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.:275 °C
Oven
Oven Temp.:120 °C (hold 1 min) to 300 °C at 10 °C/min (hold 6 min)
Carrier GasHe, constant linear velocity
Linear Velocity:65 cm/sec
DetectorECD @ 330 °C
InstrumentShimadzu 2010 GC
Notestraces shown on same scale
black trace: Restek 6mL/1g Florisil SPE Tube
red trace: competitor 6mL/1g Florisil SPE Tube