Pure Chromatography
www.restek.com
www.restek.com



This chromatogram may contain analytes that are synonyms of your search terms.
Method 8081B Chlorinated Pesticides on Rtx-CLPesticides2
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice

GC_EV1145
Column | Rtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) |
Sample | Organochlorine Pesticide Mix AB #1 (cat.# 32291) |
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000) | |
Organochlorine Pesticide Mix C #1 (cat.# 32296) | |
Diluent: | hexane |
Injection | Direct |
Liner: | Drilled Uniliner |
Inj. Temp.: | 220 °C |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 300 °C at 9 °C/min (hold 10 min) |
Carrier Gas | He |
Detector | ECD @ 310 °C |
Notes | Sample: 0.5 - 1.0 mL solvent extracted sample exchanged into hexane Standards and surrogates were spiked into hexane Tube: 3 mL, 250 mg CarboPrep 90 Tube conditioning: Apply 2 mL CH2Cl2:hexane (20:80) and pass through tube Extract cleanup: Prepare collection rack with vials, place under each tube. Add 0.5 - 1.0 mL of extracted sample to tube and collect all solutions passing through. Do not expose carbon bed to air. Add 20 mL of CH2Cl2:hexane (20:80) to tube and allow it to elute using gravity feed. Extract concentration: Solvent exchange extract to hexane Do not allow extracts to dry completely Concentrate extract to 0.5 - 1.0 mL Sample is ready for analysis. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.