Polycyclic Aromatic Hydrocarbons on Rxi-17 (15m x 0.25mm x 0.25µm)
Peaks | |
---|---|
1. | Naphthalene![]() |
2. | 2-Methylnaphthalene![]() |
3. | 1-Methylnaphthalene![]() |
4. | Acenaphthylene![]() |
5. | Acenaphthene![]() |
6. | Fluorene![]() |
7. | Phenanthrene![]() |
8. | Anthracene![]() |
9. | Fluoranthene![]() |
10. | Pyrene![]() |

GC_EV1181
Column | Rxi-17, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13520) |
Sample | |
Diluent: | Methylene chloride |
Conc.: | 20 ng/µL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 20:1) |
Liner: | 4mm Split Precision Liner w/Wool (cat.# 21022) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 42 mL/min |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 320 °C at 15 °C/min (hold 2 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 2 mL/min |
Detector | FID @ 340 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |