share:
delim
delim
Phenanthrene and Anthracene on Rxi-17 (15m x 0.25mm x 0.25µm)

Phenanthrene and Anthracene on Rxi-17 (15m x 0.25mm x 0.25µm)

Peaks
1.Phenanthrene
2.Anthracene
image
GC_EV1183
ColumnRxi-17, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13520)
Sample
Diluent:Methylene chloride
Conc.:20 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:4mm Split Precision Liner w/Wool (cat.# 21022)
Inj. Temp.:275 °C
Split Vent Flow Rate:42 mL/min
Oven
Oven Temp.:80 °C (hold 1 min) to 320 °C at 15 °C/min (hold 2 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 340 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAHs

,

50% phenyl

,

polycyclic aromatic hydrocarbons

,

polynuclear aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us