delim
delim
Peak Shape Comparison of Early Eluting Semivolatiles Using Split (10:1) and Splitless Injection at 80 °C Oven Start Temp.
This chromatogram may contain analytes that are synonyms of your search terms.

Peak Shape Comparison of Early Eluting Semivolatiles Using Split (10:1) and Splitless Injection at 80 °C Oven Start Temp.

Peaks
1.Phenol-d6 (SS)
2.Phenol
3.Aniline
4.Bis(2-chloroethyl) ether
5.2-Chlorophenol
6.1,3-Dichlorobenzene
7.1,4-Dichlorobenzene-D4 (IS)
8.1,4-Dichlorobenzene
Peaks
9.Benzyl Alcohol
10.1,2-Dichlorobenzene
11.2-Methylphenol
12.Bis(2-chloroisopropyl) ether
13.4-Methylphenol/3-Methylphenol
14.N-Nitrosodi-N-propylamine
15.Hexachloroethane
16.Nitrobenzene-D5 (SS)
17.Nitrobenzene
image
GC_EV1190
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample8270 MegaMix (cat.# 31850)
Benzoic acid (cat.# 31879)
8270 Benzidines Mix (cat.# 31852)
Acid Surrogate Mix (4/89 SOW) (cat.# 31025)
1,4-dioxane (cat.# 31853)
Revised B/N Surrogate Mix (cat.# 31887)
SV Internal Standard Mix (cat.# 31206)
Diluent:Methylene chloride
Conc.:40 µg/mL
Injectionsplitless
Inj. Temp.:270 °C
Oven
Oven Temp.:80 °C (hold 1 min) to 320 °C at 25 °C/min to 330 °C at 5 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:250 °C
Quad Temp.:150 °C
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:35-550 amu
InstrumentAgilent 7890A GC & 5975C MSD
NotesSplit: 1 μL split (10:1)
Splitless: 1 μL splitless (0.5 min.)
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us