share:
delim
delim
Organochlorine Pesticides by EPA Method 8081B on Rtx-CLPesticides/Rtx-CLPesticides2 (extended list)
This chromatogram may contain analytes that are synonyms of your search terms.

Organochlorine Pesticides by EPA Method 8081B on Rtx-CLPesticides/Rtx-CLPesticides2 (extended list)

PeaksConc.
(ng/mL)
1.1,2-Dibromo-3-chloropropane100
2.Hexachloro- cyclopentadiene100
3.Tetrachloro-m-xylene (SS)100
4.cis-Diallate500
5.Hexachlorobenzene100
6.trans-Diallate500
7.α-BHC40
8.γ-BHC40
9.β-BHC40
10.δ-BHC40
11.Heptachlor40
12.Aldrin40
13.Chlorpyrifos100
14.Isodrin100
15.Oxychlordane50
16.Heptachlor epoxide40
17.trans-Chlordane*40
18.2,4'-DDE100
19.trans-Nonachlor100
PeaksConc.
(ng/mL)
20.cis-Chlordane*40
21.Endosulfan I40
22.Captan100
23.4,4'-DDE100
24.Dieldrin80
25.2,4'-DDD100
26.Endrin80
27.Chlorobenzilate500
28.2,4'-DDT100
29.cis-Nonachlor100
30.4,4'-DDD80
31.Endosulfan II80
32.4,4'-DDT80
33.Endrin aldehyde80
34.Endosulfan sulfate80
35.Methoxychlor400
36.Mirex100
37.Endrin ketone80
38.Decachloro-biphenyl100
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV1193
ColumnsRtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261)
SampleOrganochlorine pesticide mix AB #2 (cat.# 32292)
Organochlorine pesticide mix C #1 (cat.# 32296)
2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Chlorpyrifos (cat.# 32212)
2,4'-DDE (cat.# 32099)
2,4'-DDD (cat.# 32098)
Chlorobenzilate (cat.# 32211)
Captan (cat.# custom)
cis-Nonachlor (cat.# custom)
Mirex (cat.# custom)
Oxychlordone (cat.# custom)
Diallate (cat.# custom)
Diluent:Hexane
Injection
Inj. Vol.:1 µL splitless (hold 0.3 min)
Liner:Cyclo double taper (4 mm) (cat.# 20896)
Inj. Temp.:250 °C
Oven
Oven Temp.:80 °C to 175 °C at 8 °C/min to 195 °C at 11 °C/min to 235 °C at 7 °C/min to 320 °C at 25 °C/min (hold 3 min)
Carrier GasHe, constant flow
Linear Velocity:28 cm/sec
DetectorMicro-ECD @ 330 °C
Make-up Gas Flow Rate:60 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA

,

8000

,

8081

,

Chlordane

,

BHC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us