share:
delim
delim
Pesticides and Brominated Flame Retardants by EPA Method 527 on Rxi-5Sil MS

Pesticides and Brominated Flame Retardants by EPA Method 527 on Rxi-5Sil MS

Peaks
1.1,3-Dimethyl-2-nitrobenzene
2.Acenaphthene-d10
3.Dimethoate
4.Atrazine
5.Propazine
6.Phenanthrene-d10
7.Vinclozoline
8.Prometryne
9.Bromacil
10.Malathion
11.Thiobencarb
12.Chlorpyrifos
13.Parathion
14.Terbufos sulfone
15.Oxychlordane
16.Esbiol
Peaks
17.Nitrofen
18.Kepone
19.Norflurazon
20.Hexazinone
21.Triphenyl phosphate
22.Bifenthrin
23.Chrysene-d12
24.BDE-47
25.Mirex
26.BDE-100
27.BDE-99
28.Perylene-d12
29.Fenvalerate
30.BB-153
31.Esfenvalerate
32.BDE-153
image
GC_EV1206
ColumnRxi-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620)
using IP Deactivated Guard Column 5 m, 0.53 mm ID (cat.# 10045)
with Universal Press-Tight Connectors (cat.# 20429)
SamplePBDE Mix (cat.# 33098)
Pesticides Mix #1, Method 527 (cat.# 33007)
Pesticides Mix #2, Method 527 (cat.# 33008)
Internal Standard, Method 527 (cat.# 33010)
Surrogate Standard, Method 527 (cat.# 33009)
Diluent:Ethyl acetate:methylene chloride (1:1)
Conc.:40 pg/µL (500 pg on-column)
Injection
Inj. Vol.:12.5 µL splitless (hold 0.583 min)
Liner:Gooseneck Splitless (4mm) w/Semivolatiles Wool (cat.# 20799-231.5)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:40 °C (hold 0.60 min) to 320 °C at 30 °C/min (hold 1.07 min)
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorMS
Mode:
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:-70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:3.33 min
Acquisition Range:45 - 750 amu
Spectral Acquisition Rate:10 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesCarrier Gas Flow: 2 mL/min. corrected constant flow via pressure ramps
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA 527

,

EPA method 527

,

brominated flame retardants

,

drinking water

,

pesticides

,

LVSI

,

large volume splitless injecti

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us